Product Description
The Honeywell MIDAS-E-SO2 is a gas detector sensor cartridge for sulfur dioxide (SO2). Designed for monitoring gases in laboratory or industrial environments. It features a digital display for local indication of current gas levels and peak value. The unit includes three relay outputs and a 4 to 20 mA output for integration with remote monitoring equipment.
- Three alarm levels with N.O. or N.C. alarm contacts
- 4 digit display status indication
- 4-20 mA output signal
Product Highlights
The MIDAS-E-SO2 provides a system fault relay output and a 4 to 20 mA output for integration with remote monitoring equipment or building management systems.
Product Information
Standard Information
-
Brand: HONEYWELL
-
Manufacturer Part Number: MIDAS-E-SO2
-
Country of Origin: Unknown
Technical Specifications
-
Type: Accessory
-
Supply Voltage: 24 VDC -15% to 10% @ 208 mA
-
Signal Output: 3 alarm relays, 1A @ 30 VDC or 0.5A @ 125 VAC, 4-20 mA @ 250
-
Repeatability: <± 2% of measured value
-
Linearity: <± 10% of measured value
-
Measurement Range: 0 – 8ppm
-
Gases Detected: Sulfur Dioxide (SO2)
-
Sensing Technology: 3 electrode electrochemical cell
-
Coverage: 10' (3 m) sampling and exhaust provided, 100 ft (30 m) maximum
-
Warm Up Time: < 10 minutes
-
Response Time: ≤ 5 seconds
-
Operating Temperature: 0°C to +40°C (32°F to 104°F)
-
Operating Humidity: 15 to 90% RH
-
Enclosure Rating: NEMA 1
-
Dimensions: 5.9 in H x 12.6 in W x 6 in D (15 x 6.5 x 15.3 cm)
-
Weight: 1.8 lb (0.8 Kg)
-
Approvals: ETL, CE
-
Warranty: 2 years
-
Cartridge Part Number: MIDAS-S-SO2 1 year standard warranty,
MIDAS-E-SO2 2 year extended warranty
-
Minimum Alarm 1 Set Point: 0.95ppm
-
Lower Detection Limit: 0.70ppm
-
Resolution: 0.05ppm
-
Sensor Cartridge Life Expectancy: ≥ 24 months under typical application conditions
-
Effect of Temperature Zero: <± 0.03ppm / °C
-
Effect of Temperature Sensitivity: <± 0.4% of measured value / °C
-
Effect of Humidity Zero: No effect
-
Effect of Humidity Sensitivity: <± 1% of measured value / % RH
-
Operating Pressure: 90 – 110kPa
-
Effect of Position: No effect in typical application
-
Long Term Drift Zero: TBA
-
Long Term Drift Sensitivity: <± 2% of measured value / year
-
Calibration Gas: Sulfur Dioxide (SO2)
-
Bump Test Gas: Sulfur Dioxide (SO2)
-
Storage Temperature: +5°C to +25°C (+41°F to +77°F)